Designed to achieve ultra-high stability in the temperature range of +20 °C to -40 °C

he Seta KV-2 is a two place low temperature viscosity bath designed to achieve ultra-high stability in the temperature range of +20 °C to -40 °C. The bath offers high stability of and ±0.01 at -20°C. Temperature is measured by a PT100 probe, a twin display shows the current and set temperature.

System Description

The instrument has a small footprint and is ideally suited for busy labs. It is fully self contained and requires no external chillers or connections. A heated top plate and viewing window prevent condensation and ice formation. The easy to use controller features large 13 mm digits and high intensity, long life ultra-bright LED lighting ensures clear bath visibility at all times.

  • ●  Temperature range -40°C to +20°C

  • ● High stability ±0.01 at -20°C

  • ● Accommodates up to 2 standard viscometer tubes

  • ● Illuminated Bath

  • ● Insulated and heated viewing window

  • ● Heated top plate 

Product Specialist

Mikael Alfredsson Kovalent

Mikael Alfredsson

"Our expert in Digestion, Extraction, Syntesis and Evaporation with more than 20 years experience in the field of sales and service. Contact me for if you have questions or need a quotation"

Additional Information

The KV2 bath requires 2 Ubbelodhe Viscometer tubes. The maximum viscosity of jet fuel at -20 °C is 8 mm2/s so we recommend that one each of the following viscometers are used:

• Calibrated Ubbelodhe Viscometer size code 0B which has a viscosity range of 1 to 5 mm2/s. Seta Ref 11643/03
• Calibrated Ubbelodhe Viscometer size code 1 which has a viscosity range of 2 to 10 mm2/s. Seta Ref 11643/04

The bath has an independent over-temperature and low fluid level cut-out and a warning alarm for safety in the laboratory environment.

Data logging via the external RS-232 port is possible when connected to a PC and suitable software.

The insulated bath is sealed by 8 mm thick glass front and rear with 60 mm plexiglass providing additional insulation.

Supplied with:
2 covers for test wells
Mains lead
Instruction manual

Excludes:
Viscometer Tubes

Industries

  • ● Lubricants

  • ● Petroleum - Fuels & Oils

  • ● Chemicals & Solvents

Methods

  • ● ASTM D445

  • ● ASTM D446

  • ● IP 71

  • ● BS 188

  • ● BS 2000-71-2 & ISO 3105

  • ● BS EN ISO 3104 & BS 2000-71.1 & ISO 3104